I.Summary
The instrument is designed and made as per industrial standard of People's Republic of China SH/T0193-92 Lubricating oils-Determination of oxidation stability-Rotating pressure vessel method and ASTM D2272.
The instrument is used to determine the oxidation stability of steam turbine with the same composition (oil base oil and additive) .Also can be used to determine new mineral insulating oil containing 2,6-BHT.
II. Main technical specifications and parameters
1. Power supply: AC220V±10%
2. Heating tube power: 2500W
3. Range of pressure sensor:(0~1.6)MPa,accuracy:±2‰
4.Temperature controlling range of oil bath:room temperature~200.0℃, adjustable continuously.150.0℃ in common use.
5.Temperature controlling accuracy:±0.1℃
6.Test sample: Two-bomb design, can do two samples at one time. Convenient to do parallel test.
7. Rotation speed: (100±5)r/min
8. Included angle between oxygen bomb and water level: 30°
9. Cubage of oil bath: 30L